Quick Contact>

Hotline

0731-85260926 18665876229

中文
Power Transistor Curve Tracer Semiconductor Parameter Analyzer SMU Source Measure Unit Dynamic Parameter Test System CV Parameter Tester Gate Charge (Qg) Tester WAT Wafer Acceptance Test System TLP Transmission Line Pulse EMMI Low-Light Infrared Fault Localization HCI/BTI/TDDB/EM effects Femtoammeter/Electrometer/High Resistance Meter High and low temperature electrical performance simulation H3TRx Customizable System Integration Software
Standard Manual Probe Station Portable Desktop Probe Station Semi-automatic probe station RF/Microwave Probe Station High/Low Temperature Probe Station Vacuum cryogenic probe station system Wafer electrical parameter test system Strong Magnetic - Hall Probe Station Custom Precision Probe Card High/Low Temperature Chuck System (Taiko) Magnetic Probe Holder CCD Microscope Test cables and adapter fixtures Shielded Box and Vibration Isolation Table Probes and other accessories
B-H Analyzer Magnetic Loss Power Tester Impedance Analyzer DC Bias High-Frequency LCR Meter Vector Network Analyzer VNA Frequency Response Analyzer Milliohm Meter / Battery Internal Resistance Tester
Rapid Temperature Change Thermal Flow Meter Device High/Low Temperature Platform Thermalplate Device junction temperature simulator Tj
12-bit high-definition digital oscilloscope Optically Isolated Oscilloscope System Basic Digital Oscilloscope Signal generator Digital Multimeter DMM Spectrum Analyzer Frequency counter Infrared thermal imager
Optically isolated voltage probe Optically Isolated Current Probe Dynamic On-Resistance Rds(on)dynamic Oscilloscope voltage probe High-voltage differential probe Fluxgate current transformer AC/DC Current Probe Non-inductive coaxial shunt Communication Interfaces and Cables Dedicated fixture Software
Multi-channel data acquisition instrument Data Acquisition and Playback Power Analyzer
Digital Phase-Locked Loop Amplifier Ultra-Low Noise Preamplifier Precision Current Amplifier Precision Differential Amplifier Precision Filter Light chopper Precision Low-Noise DC Power Supply
Programmable DC Power Supply DC Electronic Load Precision Bipolar Power Amplifier Bipolar simulation power supply
Safety Tester (ACI/DCI/IR/GB) High-voltage meter HV meter
Standard or calibration source meter Standard current and voltage meters
Products
Lianxun Instruments S0342C 4-channel PXIe plug-in source meter precision power supply/measurement unit ±30 V, ±500 mA (DC/pulse)

Lianxun Instruments S0342C 4-channel PXIe plug-in source meter precision power supply/measurement unit ±30 V, ±500 mA (DC/pulse)

Model:S0342C      Country of Origin:China

Category:SMU Source Measure Unit

Overview:

Semight Instruments S0342C is a compact, cost-effective four-channel PXIe source/measure unit (SMU) capable of sourcing and measuring voltage and current simultaneously, with a maximum output of ±30 V and ±500 mA (DC/pulse). It supports conventional SMU SCPI commands for quick and easy migration of existing test code, is compatible with mainstream PXIe chassis, and supports multi-card synchronization for integration into production test systems, improving test throughput and reducing cost.



Features
Specifications
Application Notes
Functions & Configuration

Key Features

  • Easy to expand

    4-channel standard PXIe precision power supply, easily scalable
  • Wide range

    Ranges: ±30 V, ±500 mA (DC/pulsed)
  • High resolution

    Minimum measurement resolution down to 100 pA / 60 μV
  • High sampling rate

    Supports ADC sampling rates up to 500 K

Features and Benefits

  • 5 functions in one

    Voltage source
    Current source
    Ammeter
    Voltmeter
    Electronic load
  • Quadrants I and III operate as a source: the actual polarity of the V/I output follows the source setting;
    Quadrants II and IV operate as a load: CC and CV work together; when used as a load, the load setting polarity is opposite to the source polarity;
  • Capable of testing a wide range of devices

  • Capture more measurement data

    ♦ 6-digit resolution: accuracy comparable to a 6½-digit multimeter;
    ♦ 100 pA / 60 µV resolution: excellent measurement sensitivity;
    ♦ 500 kSa/s sampling rate: high-speed measurement with fast setup/digitizing rate for arbitrary waveform generation (list sweep);
  • Extensive sweep capabilities


Voltage specifications



Voltage setting accuracy
RangeSetting resolutionAccuracy (1 year)1±(% reading + offset)Typical noise (RMS) 0.1 Hz-10 Hz
±30 V

1 mV

0.03%+4 mV1000 μV
±6 V

200 μV

0.03%+1 mV100 μV
Temperature coefficient±(0.15 × accuracy specification)/°C (0℃-18℃, 28℃-50℃)
Channel 23CH0 to CH3
Output power

Maximum 3 W per channel, maximum 6 W total for four channels

settling time

<200 μs (typical)

Overshoot<±0.1% (typical, Normal, step is 10% to 90% of range, full-scale point, resistive load test)
Noise 10Hz-20MHz

6 V voltage source, 0.5 A resistive load, <3 mVrms

[1] Accuracy calculation example: to test the accuracy of a 1 V output on the 6 V range, the tolerance is:

[2] All channel outputs are electrically isolated from earth ground, but the outputs of all channels share a common ground (LO)

[3] This instrument outputs potentially hazardous high voltage (±31.5 V) to the HI/Sense HI/Guard terminals. To prevent electric shock, relevant safety precautions must be taken before power-on. Do not connect the Guard terminal to any output, including shorting it to chassis ground or output LO, otherwise the instrument will be damaged.



I-V output capability





Current source specifications

Current accuracy

Range

Setting resolution

Accuracy (1 year)

±(% reading + offset)

Typical noise (RMS)

0.1Hz-10Hz

±500 mA【4】

20 μA

0.05% + 100 μA + Vo*25 μA

10 μA

±100 mA

4 μA

0.05% + 10 μA+ Vo*5 μA

1 μA

±10 mA

400 nA

0.05% + 5 μA + Vo*500 nA

100 nA

±1 mA

40 nA

0.05% + 500 nA + Vo*50 nA

10 nA

±100 μA

4 nA

0.05% + 50 nA + Vo*5 nA

1 nA

±10 μA

400 pA

0.05% + 20 nA + Vo*500 pA

150 pA

Temperature coefficient

±(0.15×accuracy specification)/℃ (0℃-18℃, 28℃-50℃)

Channel

CH0 to CH3

Output power

Max. 3 W per channel, max. 6 W total for four channels

Overshoot

<±0.1% (typical, Normal, step from 10% to 90% of range, full-scale point, resistive load test)

[4] The 500 mA range supports the 6 V voltage range only.



Voltmeter specifications

Voltmeter accuracy

Range

Measurement resolution

Accuracy (1 year)

±(% reading + offset)

±30 V

300 μV

0.03%+4 mV

±6 V

60 μV

0.03%+1 mV

Temperature coefficient

±(0.15×accuracy specification)/℃ (0℃-18℃, 28℃-50℃)



Ammeter specifications

Ammeter accuracy

Range

Measurement resolution

Accuracy (1 year)

±(% reading + offset)

±500 mA【4】

10 μA

0.05% + 100 μA + Vo*25 μA

±100 mA

1 μA

0.05% + 10 μA+ Vo*5 μA

±10 mA

100 nA

0.05% + 5 μA + Vo*500 nA

±1 mA

10 nA

0.05% + 500 nA + Vo*50 nA

±100μA

1 nA

0.05% + 50 nA + Vo*5 nA

±10μA

100 pA

0.05% + 20 nA + Vo*500 pA

Temperature coefficient

±(0.15×accuracy specification)/℃ (0℃-18℃, 28℃-50℃)