Quick Contact>

Hotline

0731-85260926 18665876229

中文
Power Transistor Curve Tracer Semiconductor Parameter Analyzer SMU Source Measure Unit Dynamic Parameter Test System CV Parameter Tester Gate Charge (Qg) Tester WAT Wafer Acceptance Test System TLP Transmission Line Pulse EMMI Low-Light Infrared Fault Localization HCI/BTI/TDDB/EM effects Femtoammeter/Electrometer/High Resistance Meter High and low temperature electrical performance simulation H3TRx Customizable System Integration Software
Standard Manual Probe Station Portable Desktop Probe Station Semi-automatic probe station RF/Microwave Probe Station High/Low Temperature Probe Station Vacuum cryogenic probe station system Wafer electrical parameter test system Strong Magnetic - Hall Probe Station Custom Precision Probe Card High/Low Temperature Chuck System (Taiko) Magnetic Probe Holder CCD Microscope Test cables and adapter fixtures Shielded Box and Vibration Isolation Table Probes and other accessories
B-H Analyzer Magnetic Loss Power Tester Impedance Analyzer DC Bias High-Frequency LCR Meter Vector Network Analyzer VNA Frequency Response Analyzer Milliohm Meter / Battery Internal Resistance Tester
Rapid Temperature Change Thermal Flow Meter Device High/Low Temperature Platform Thermalplate Device junction temperature simulator Tj
12-bit high-definition digital oscilloscope Optically Isolated Oscilloscope System Basic Digital Oscilloscope Signal generator Digital Multimeter DMM Spectrum Analyzer Frequency counter Infrared thermal imager
Optically isolated voltage probe Optically Isolated Current Probe Dynamic On-Resistance Rds(on)dynamic Oscilloscope voltage probe High-voltage differential probe Fluxgate current transformer AC/DC Current Probe Non-inductive coaxial shunt Communication Interfaces and Cables Dedicated fixture Software
Multi-channel data acquisition instrument Data Acquisition and Playback Power Analyzer
Digital Phase-Locked Loop Amplifier Ultra-Low Noise Preamplifier Precision Current Amplifier Precision Differential Amplifier Precision Filter Light chopper Precision Low-Noise DC Power Supply
Programmable DC Power Supply DC Electronic Load Precision Bipolar Power Amplifier Bipolar simulation power supply
Safety Tester (ACI/DCI/IR/GB) High-voltage meter HV meter
Standard or calibration source meter Standard current and voltage meters

Technical Services & Solutions

Technical support and after sales service

Technical Services & Solutions

What is the measurement range of magnetic flux density B and magnetic field strength H for the Iwasaki BH analyzer?

Published: 2024-01-02 Views: 次

BH.jpg

Q. What is the flux density range that can be measured with a B-H analyzer?

A. Under sinusoidal excitation in the non-saturated region, the relationship between the maximum flux density Bm of the B-H curve and the induced voltage V2 is given by the following formula.

f: measurement frequency Ae: effective cross-sectional area of the specimen N2: number of turns of the secondary winding
formula
Therefore, measurement can be performed over the Bm range corresponding to V2 within the voltage detection range of the B-H analyzer used in the non-saturated region. The maximum allowable V2 for the SY-8218 and SY-8219 is 200 V. Note that this formula does not hold under non-sinusoidal excitation or in the saturated region.

Q. What magnetic field (strength) can be excited by a B-H analyzer?

A. The maximum magnetic field Hm that can be excited is determined by the following formula.

I: excitation current N1: number of turns of the primary winding Le: mean magnetic path length
formula
The excitation current depends on the power amplifier used.

bh_6.jpg