Quick Contact>

Hotline

0731-85260926 18665876229

中文
Power Transistor Curve Tracer Semiconductor Parameter Analyzer SMU Source Measure Unit Dynamic Parameter Test System CV Parameter Tester Gate Charge (Qg) Tester WAT Wafer Acceptance Test System TLP Transmission Line Pulse EMMI Low-Light Infrared Fault Localization HCI/BTI/TDDB/EM effects Femtoammeter/Electrometer/High Resistance Meter High and low temperature electrical performance simulation H3TRx Customizable System Integration Software
Standard Manual Probe Station Portable Desktop Probe Station Semi-automatic probe station RF/Microwave Probe Station High/Low Temperature Probe Station Vacuum cryogenic probe station system Wafer electrical parameter test system Strong Magnetic - Hall Probe Station Custom Precision Probe Card High/Low Temperature Chuck System (Taiko) Magnetic Probe Holder CCD Microscope Test cables and adapter fixtures Shielded Box and Vibration Isolation Table Probes and other accessories
B-H Analyzer Magnetic Loss Power Tester Impedance Analyzer DC Bias High-Frequency LCR Meter Vector Network Analyzer VNA Frequency Response Analyzer Milliohm Meter / Battery Internal Resistance Tester
Rapid Temperature Change Thermal Flow Meter Device High/Low Temperature Platform Thermalplate Device junction temperature simulator Tj
12-bit high-definition digital oscilloscope Optically Isolated Oscilloscope System Basic Digital Oscilloscope Signal generator Digital Multimeter DMM Spectrum Analyzer Frequency counter Infrared thermal imager
Optically isolated voltage probe Optically Isolated Current Probe Dynamic On-Resistance Rds(on)dynamic Oscilloscope voltage probe High-voltage differential probe Fluxgate current transformer AC/DC Current Probe Non-inductive coaxial shunt Communication Interfaces and Cables Dedicated fixture Software
Multi-channel data acquisition instrument Data Acquisition and Playback Power Analyzer
Digital Phase-Locked Loop Amplifier Ultra-Low Noise Preamplifier Precision Current Amplifier Precision Differential Amplifier Precision Filter Light chopper Precision Low-Noise DC Power Supply
Programmable DC Power Supply DC Electronic Load Precision Bipolar Power Amplifier Bipolar simulation power supply
Safety Tester (ACI/DCI/IR/GB) High-voltage meter HV meter
Standard or calibration source meter Standard current and voltage meters

Technical Services & Solutions

Technical support and after sales service

Technical Services & Solutions

What are the application advantages of Iwasaki B-H analysis? How should models be selected?

Published: 2023-12-29 Views: 次

With their exceptionally wide test frequency bandwidth, outstanding accuracy and stability, and highly flexible, future-proof design, IWATSU BH analyzers are the unanimous choice of software R&D professionals worldwide.

BH.jpg


Q. What frequency range do IWATSU B-H analyzers measure?


A. This depends on the options configured with the B-H analyzer. The measurement frequency of each standalone B-H analyzer model is shown in the table below. For measurement frequencies with options installed, refer to the documentation for each option.

ModelMeasurement frequency
SY-821810 Hz to 10 MHz
SY-821910 Hz to 1 MHz


In terms of performance, the B-H analyzer can measure across the frequency ranges above, but at the upper limit, pay attention to the inductance L of the device under test (DUT).

The input capacitance C of the B-H analyzer and the inductance L of the DUT form an LC parallel resonant circuit, so the measurement frequency must be kept below 1/10 of the resonant frequency. Please take this into account when selecting a model. The resonant frequency fc can be calculated using the following formula.
L: DUT inductance C: input capacitance of the measuring instrument (approx. 18.5 pF for SY-8218 / SY-8219) N1: number of primary winding turns N2: number of secondary winding turns

formula

In addition, for DUTs with a high phase angle (≧ approx. 87°), resonance effects become more pronounced; when measuring such samples, we recommend measuring at 1/20 or less of the resonant frequency.

Q. Can DC bias measurements be performed with a B-H analyzer?


A. Yes. Combine the SY-8218 / SY-8219 with the optional DC bias test unit SY-960 (SY-961, SY-962). There is no need to wind a dedicated third winding on the DUT for DC excitation, making measurement far simpler than with conventional products.


Q. Can pulse excitation measurements be performed with a B-H analyzer?


A. The SY-8218 / SY-8219 support pulse excitation measurements within the standard 10 Hz to 1 MHz range, limited to 50% duty. A. When combined with the optional DC bias test unit SY-960 (SY-961, SY-962), measurements can be performed over 10 kHz to 1 MHz with an adjustable duty of 10% to 90%.